Молимо вас користите овај идентификатор за цитирање или овај линк до ове ставке:
https://open.uns.ac.rs/handle/123456789/8683
Назив: | Monte Carlo simulation of GM probe and NaI detector efficiency for surface activity measurements | Аутори: | Grujić, Selena Djordjević I. Milošević, Miodrag Kozmidis-Luburić, Uranija |
Датум издавања: | 5-сеп-2013 | Часопис: | Radiation Measurements | Сажетак: | This paper deals with the direct measurement of total (fixed plus removable) surface activity in the presence of interfering radiation fields. Two methods based on Monte Carlo simulations are used: one for a Geiger-Muller (GM) ionisation probe and the other for sodium iodide (NaI) detector with lead collimators; equations for the most general case and the geometry models for Monte Carlo simulation of both (GM and NaI) detectors are employed. Finally, an example of application is discussed. © 2013 Elsevier Ltd. All rights reserved. | URI: | https://open.uns.ac.rs/handle/123456789/8683 | ISSN: | 13504487 | DOI: | 10.1016/j.radmeas.2013.08.002 |