Mоlimо vаs kоristitе оvај idеntifikаtоr zа citirаnjе ili оvај link dо оvе stаvkе: https://open.uns.ac.rs/handle/123456789/14108
Nаziv: Further generalization of the low-frequency true-RMS instrument
Аutоri: Pjevalica V.
Vujičić V.
Dаtum izdаvаnjа: 1-мар-2010
Čаsоpis: IEEE Transactions on Instrumentation and Measurement
Sažetak: This paper shows that the use of random uniform dither in harmonic measurement can significantly shorten both the word of the input A/D converter and the word of the applied base function (sine and/or cosine, prestored in memory), without accuracy loss. This simplifies the hardware for a precise instrument for harmonic measurement. Theoretical considerations demonstrate that the upper limit of the absolute measurement uncertainty is identical for every harmonic coefficient. It is shown theoretically, by simulation, and by experiment that a 6-bit dithered A/D converter word, an 8-bit dithered base function word, and a 6 × 8 (14-bit) multiplier word assure a measurement of 16 harmonics (16 sine and 16 cosine components) with 13-bit accuracy. The measurement of 1 × 16 harmonics is realized in a small programmable-logic-device (PLD) chip. © 2009 IEEE.
URI: https://open.uns.ac.rs/handle/123456789/14108
ISSN: 00189456
DOI: 10.1109/TIM.2009.2030874
Nаlаzi sе u kоlеkciјаmа:FTN Publikacije/Publications

Prikаzаti cеlоkupаn zаpis stаvki

SCOPUSTM   
Nаvоđеnjа

28
prоvеrеnо 10.05.2024.

Prеglеd/i stаnicа

4
Prоtеklа nеdеljа
2
Prоtеkli mеsеc
0
prоvеrеnо 10.05.2024.

Google ScholarTM

Prоvеritе

Аlt mеtrikа


Stаvkе nа DSpace-u su zаštićеnе аutоrskim prаvimа, sа svim prаvimа zаdržаnim, оsim аkо nije drugačije naznačeno.