Mоlimо vаs kоristitе оvај idеntifikаtоr zа citirаnjе ili оvај link dо оvе stаvkе: https://open.uns.ac.rs/handle/123456789/13900
Nаziv: Multi-field surface electrode for selective electrical stimulation
Аutоri: Popović-Bijelić A.
Bijelić G.
Jorgovanović, Nikola 
Bojanić, Dubravka 
Popović M.
Popović D.
Dаtum izdаvаnjа: 1-јан-2005
Čаsоpis: Artificial Organs
Sažetak: We designed a 24-field array and an on-line control box that selects which and how many of 24 fields will conduct electrical charge during functional electrical stimulation. The array was made using a conductive microfiber textile, silver two-component adhesive, and the conductive ink imprint on the polycarbonate. The control box comprised 24 switches that corresponded one-to-one to the fields on the array. Each field could be made conductive or nonconductive by simple pressing of the corresponding push-button type switch on the control box. We present here representative results of the selectivity of the new electrode measured in three tetraplegic patients during functional electrical stimulation of the forearm. The task was to generate finger flexion and extension with minimal interference of the wrist movement during lateral and palmar grasps. Therapists determined the appropriate pattern that lead to effective grasping, lasting on average 5 min per stimulation channel in the first session. This optimal conductive pattern (size and shape) provided effective finger flexion and extension with minimal wrist flexion/extension and ulnar/radial deviations (<10 degrees). The optimal size and shape of the electrode in all cases had a branched pattern. The selection of the optimal stimulation site was achieved without moving the electrode. The size and shape were reproducible in the same subject from session to session, yet were different from subject to subject. The optimal electrode size and shape changed when subjects pronated and supinated their forearm. The control box includes a program that can dynamically change the number and sites of the conductive fields; hence, it is feasible to use this during functional movements. Subjects learned how to determine the optimal electrode pattern; hence, these electrodes could be effective for home usage. © 2005 International Center for Artificial Organs and Transplantation.
URI: https://open.uns.ac.rs/handle/123456789/13900
ISSN: 0160564X
DOI: 10.1111/j.1525-1594.2005.29075.x
Nаlаzi sе u kоlеkciјаmа:FTN Publikacije/Publications

Prikаzаti cеlоkupаn zаpis stаvki

SCOPUSTM   
Nаvоđеnjа

68
prоvеrеnо 10.05.2024.

Prеglеd/i stаnicа

21
Prоtеklа nеdеljа
11
Prоtеkli mеsеc
2
prоvеrеnо 10.05.2024.

Google ScholarTM

Prоvеritе

Аlt mеtrikа


Stаvkе nа DSpace-u su zаštićеnе аutоrskim prаvimа, sа svim prаvimа zаdržаnim, оsim аkо nije drugačije naznačeno.