Please use this identifier to cite or link to this item: https://open.uns.ac.rs/handle/123456789/12251
Title: Cost-effective microfluidic device for detection of psychoactive substances
Authors: Laketa J.
Kojić T.
Radovanović M.
Kojić S.
Milićević, Dragan 
Stojanović, Goran 
Issue Date: 7-Feb-2019
Journal: 2018 International Symposium on Industrial Electronics, INDEL 2018 - Proceedings
Abstract: © 2018 IEEE. This paper presents design and fabrication of cost-effective microfluidic chip using xurographic technique. Seven layers of PVC foils and gold electrodes as a plane of microfluidic device were laminated in order to obtain a compact chip. Syringe pump was used for introducing fluid in one inlet and when fluid went through the microchannel the electrical parameters of capacitive structure has been changed which was measured by means of Impedance analyser. Using this method, it is possible to detect concentration of psychoactive substances in low volume in aqueous or biological media (such as saliva), very efficiently.
URI: https://open.uns.ac.rs/handle/123456789/12251
ISBN: 9781538623534
DOI: 10.1109/INDEL.2018.8637618
Appears in Collections:FTN Publikacije/Publications

Show full item record

SCOPUSTM   
Citations

1
checked on Sep 14, 2022

Page view(s)

21
Last Week
0
Last month
0
checked on May 10, 2024

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.