Назив: Improvement of measuring methods and instrumentation concerning <sup>222</sup>Rn determination in drinking waters - RAD7 and LSC technique comparison
Аутори: Stojković Ivana 
Tenjović Branislava
Nikolov Jovana 
Vesković Miroslav 
Mrđa Dušan 
Todorović Nataša 
Датум издавања: 2015
Часопис: Applied Radiation and Isotopes
Сажетак: © 2015 Elsevier Ltd. A procedure for the determination of 222Rn in environmental water samples using liquid scintillation counting (LSC) was applied and optimized. A minimum detectable activity of 0.029Bql-1 in a 20ml glass vial (10ml water sample mixed with 10ml of liquid scintillation cocktail) has been achieved during 300min of measurement time. The procedure was compared with RAD7 radon detector measurements. 226Ra content in the water was determined by gamma-ray spectroscopy. Applications to drinking waters collected from public drinking fountains in the Vojvodina (Serbia) are presented with annual effective dose for ingestion and inhalation for adults calculated.