Назив: Is attenuation technique sensitive enough to be used to make a distinction between flattening filter and flattening filter free spectrum of a therapy linac
Аутори: Toth Arpad A.
Dragojlovi Luka D.
Ignjatov Nevena V.
Mikalački Mirjana
Krmar M. 
Датум издавања: 2019
Часопис: Nuclear Technology and Radiation Protection
Сажетак: © 2019, Vinca Inst Nuclear Sci. All rights reserved. The purpose of this study was to check if attenuation analysis is sensitive enough to make a distinction between photon energy spectra of standard accelerators and accelerators operating without a field flattening filter. Attenuation measurements were performed in two operating modes of a medical linear accelerator: with a flattening filter and without it, at nominal energy of 6 MV. Numerical and Laplace transform pair methods were used for spectra reconstruction. Although the difference between two attenuation curves is very small, as expected both reconstruction techniques produce slightly softer spectra at flattening filter free mode. The most important conclusion of this pilot study is that attenuation analysis can be a sensitive tool which is able to detect spectral differences between photon beams of an accelerator operating in two regimes.
URI: https://open.uns.ac.rs/handle/123456789/20099
ISSN: 1451-3994
DOI: 10.2298/NTRP181016010T
Налази се у колекцијама:PMF Publikacije/Publications

Приказати целокупан запис ставки

Преглед/и станица

4
Протекла недеља
2
Протекли месец
0
проверено 10.05.2024.

Google ScholarTM

Проверите

Алт метрика


Ставке на DSpace-у су заштићене ауторским правима, са свим правима задржаним, осим ако није другачије назначено.