Mоlimо vаs kоristitе оvај idеntifikаtоr zа citirаnjе ili оvај link dо оvе stаvkе:
https://open.uns.ac.rs/handle/123456789/12291
Nаziv: | Measurement of complex permeability using short coaxial line reflection method | Аutоri: | Radonić, Vasa Blaž, Nelu Živanov, Ljiljana |
Dаtum izdаvаnjа: | 2010 | Izdаvаč: | Polish Academy of Sciences | Čаsоpis: | Acta Physica Polonica A | Sažetak: | This paper describes realization of complex permeability determination for ferrite materials using short coaxial sample holder and vector network analyzer E5071B in frequency range between 300 kHz and 1 GHz. The design of coaxial high frequency sample holder is presented, the principle of measurement and calibration is shown. The extracting of complex permeability from S-parameters is managed based on accomplished formulas, user-friendly program for computer control has been written, and obtained results are explained in detail. In order to verify proposed method, the results of the measurement of NiZn ferrite samples, manufactured by MMG NeoSid, are compared with catalog characteristics. | URI: | https://open.uns.ac.rs/handle/123456789/12291 | ISSN: | 0587-4246 | DOI: | 10.12693/APhysPolA.117.820 |
Nаlаzi sе u kоlеkciјаmа: | IBS Publikacije/Publications |
Prikаzаti cеlоkupаn zаpis stаvki
SCOPUSTM
Nаvоđеnjа
10
prоvеrеnо 14.09.2022.
Prеglеd/i stаnicа
33
Prоtеklа nеdеljа
17
17
Prоtеkli mеsеc
0
0
prоvеrеnо 03.05.2024.
Google ScholarTM
Prоvеritе
Аlt mеtrikа
Stаvkе nа DSpace-u su zаštićеnе аutоrskim prаvimа, sа svim prаvimа zаdržаnim, оsim аkо nije drugačije naznačeno.