Title: | Electrochemical Detection of MMP-2 Using Graphene-Based Aptasensor | Authors: | Jarić, Stefan Schobesberger, Silvia Ertl, Peter Knežević, Nikola Bobrinetskiy, Ivan |
Keywords: | MMP2; graphene; aptamer; biosensor | Issue Date: | Mar-2024 | Conference: | XXXV Eurosensors Conference, Lecce, Italy, September 10–13, 2023 | Abstract: | A graphene-based electrochemical biosensor was developed for the detection of matrix metalloproteinase 2 (MMP-2) endopeptidase, whose expression can be significantly related to the occurrence, metastasis, and prognosis of cancer. A specific anti-MMP-2 aptamer was successfully immobilized on the surface of electrochemically reduced graphene oxide via a pyrene-based linker, enabling the specific capture of MMP-2. The sensor was able to detect 1 ng mL−1, with an overall detection time of less than 20 min. Moreover, the aptamer-based biosensor showed good specificity toward different unspecific proteins. | URI: | https://open.uns.ac.rs/handle/123456789/32724 | DOI: | 10.3390/proceedings2024097057 |
Appears in Collections: | IBS Publikacije/Publications |
Files in This Item:
File | Size | Format | |
---|---|---|---|
proceedings-97-00057.pdf | 463.32 kB | Adobe PDF | View/Open |